WARNING
Always observe the following when installing, operating or maintaining
this product:
This unit must be connected to a power distribution system that has a
direct connection to earth ground (Terminated Terra [TT] network/
ground connected). This unit is not suitable for use on a floating
ground (Interrupted Terra [IT]) network.
The AC input power cord must be shielded and must have a minimum
current rating of 10 A with a nominal cross-section area of 0.75 sq mm
(reference: AWG #18) per conductor, 2 wires plus ground and product
safety approvals as required for use in the country in which the unit is
installed.
When the unit is mounted in an equipment rack or cabinet, be certain
that the internal temperature within the rack or cabinet does not
exceed the limits defined in the Product Specification or this
document.
To ensure the integrity of safety features of this unit, maintenance must
be performed only by qualified service personnel using designated
Plasmon LMS parts.
In case of fire or other emergency, isolate the units from the main
power by disconnecting the power plugs from their site power
receptacles.  In situations where disconnecting the plugs is not
possible or practical, use the system main power disconnect to isolate
the units from the main power.
To prevent fire or shock hazard,  do not expose this unit to rain or
moisture. Refer servicing to qualified technicians.