SECTION 5—OPERATING INSTRUCTIONS
21
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm‐up 
time (approximately 30 min.) while waiting for the O
2
 purity to 
reach maximum.
Front Panel Indicators - Units without 
SensO
2
NOTE: For this procedure, refer to FIGURE 5.7.
LABEL 
SYMBOL
O
2
 PURITY
INDICATOR LIGHTS (LED)
SYSTEM OKAY
O
2
 over 85% (± 2%)
GREEN Indicator Light
O
2
 Between 73% (± 3%) 
to 85% (±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor 
Failure
Call a qualified technician. Have a 
back-up supply of oxygen ready.
SYSTEM FAILURE 
O
2
 Below 73% (±3%)
RED Indicator Light
Continuous Audible Alarm 
Sieve-GARD Compressor Shut-
down. Call a qualified technician. 
Switch to a back-up supply of oxy-
gen immediately.
O
2
I/O
Auto Shut Down 
Use Backup
Normal
RED
GREEN
Call Supplier