Safety Precautions 
AC Power Supply 
The HGMSZ uses a universal power supply that is capable of accepting inputs of 100 to 240 VAC, 
50/60 Hz. The monitor’s power consumption is 15 Watts. It is highly suggested that the monitor be 
connected directly to the AC power source, preferably on its own circuit (with UPS or surge protection).  
A switch or circuit breaker rated 1.0 A, 250 VAC, with a minimum terminal spacing of 3.0 mm must be 
attached to the monitor’s AC power leads. This switch must also be located in close proximity to the 
monitor, and be in easy reach of the operator. This switch should also be clearly marked as the monitor’s 
main AC disconnect device. 
Protective Grounding 
Under no circumstances should the HGMSZ be operated without connection to a protective ground. Doing 
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this 
type of equipment.  
Explosive Atmosphere 
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any 
electrical equipment in such an environment constitutes a safety hazard. 
Proper Exhaust Venting 
It is imperative that the monitor’s exhaust port be vented as described in this manual. Failure to do so 
may constitute a safety hazard. 
Working Inside Monitor 
Extreme care should be exercised when accessing the interior of the monitor. Only qualified electrical 
maintenance personnel should perform connections and adjustments. Always remove AC power before 
working inside the monitor. 
Misuse and Modifications to Monitor 
The protection provided by the monitor may be impaired if the monitor is used in a manner not specified by 
Bacharach, Inc. Changes or modifications to this monitor, not expressly approved, will void the warranty. 
In Case of Malfunction 
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of 
such occurrence, de-energize the power supply and contact a qualified repair technician or the nearest 
Bacharach Service Center. Use ONLY the provided knockouts for electrical and communication wiring.  
Drilling into the box will void the warranty.  
Fusing 
F1, F2: 1.0 A, 250 V, Type “F” 
Installation Category 
Installation Category II, Pollution Degree II, as defined by UL. 
Altitude Limit 
6,562 ft (2,000 m) 
Cleaning 
To clean the outside of the case, DO NOT use soap and water. USE a dry cloth.