2.5  Configuring the SETUPS Function 
There are four user-defined Setups that allow the operator to quickly switch between frequently used 
combinations of Gas Type, Range, Feedback Mode, Setpoint, and the Auto Zero Off condition. Program the 
four Setup configurations as follows: 
 
First enter the Setups menu by pressing both the ENT and ESC buttons on either the instrument’s front 
panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to 
highlight “SETUPS”, and then press ENT to display the Setups menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
 
 
NOTE: 
 The names listed in the Setups menu are user defined, and may vary from what is shown above. 
 
From the Setups menu, use the Up and Down buttons to highlight the desired Setup. Press ENT to select 
that Setup and display its Configuration menu. 
 
From the Configuration menu, use the Up and Down buttons to highlight the desired function. Press 
ENT to select that function and display its setup screen. Perform the appropriate setup procedure as 
2.5.1 Setup Name 
 
 
 
 
The SETUP NAME function allows the operator to give each Setup a unique name that will appear on the 
Data Display screen. Use the Left and Right buttons to move the blinking cursor across the screen to 
highlight the character to be modified, and then use the Up and Down buttons to modify that character. 
When finished entering the desired characters, press ENT to save the new name. 
 
2.5.2 Range 
 
 
 
 
The RANGE function sets the sensitivity of the gas-level graphical display (bar graph located at the 
bottom of the display), audible tone frequency, and the hand-held probe’s LED blink rate in relation to the 
detected gas level. Use the Up and Down buttons to highlight the desired sensitivity. Press ENT to save 
the selection. 
NOTE: 
The range setting has no affect on the accuracy of the gas reading as displayed on the instrument 
when making a leak-rate measurement.  
 
SETUP NAME #1 
 
        
S
ETUP NO. 1 
Setups Menu:  
 
SETUP NO.1 
 SETUP 
NO.2 
 SETUP 
NO.3 
 SETUP 
NO.4 
Configuration Menu: 
    SETUP NAME 
    RANGE 
    GAS TYPE 
    FEEDBACK MODE 
    SETPOINT 
    SEARCH AUTO ZERO 
RANGE LG 
SETUP NO. 1 
MED 
 
SM