Bacharach H25-IR User Manual

Page of 60
H25-IR 
Calibration 
 
Instruction 3015-4342 
23
 
3.2 Internal Calibration 
Internal calibration uses the instrument’s built-in leak source. Properties of the leak source (i.e. gas type, 
leak rate, temperature coefficient, and specification temperature) are entered at the factory and do not 
need to be changed by the operator. 
The properties of the leak source along with the current internal temperature of the instrument are used 
to compute the actual leak rate when a calibration is requested. This number is displayed as the “target” 
value during calibration. A spring switch located in front of the CAL Port ensures that the probe is kept in 
place for the duration of the calibration process. Calibration will fail if the probe is withdrawn too early in 
the process, or if the computed parameters deviate from the stored factory calibration by more than 20%. 
1.  Turn ON instrument and allow it to complete its 120 second warm-up. 
2.  Press the front panel CAL button to display the Calibration screen. Note that the display must be 
reading a leak rate of “0.00” when the CAL button is pressed; otherwise, the message “LEAK 
RATE MUST BE ZERO TO CALIBRATE” will appear. If necessary, press the ZERO button to 
zero the display before pressing CAL
 
 
 
 
3.  Ensure that the instrument is set up for internal calibration as indicated by “INT CAL” being 
displayed at the bottom-right of the display. If “EXT CAL” (external calibration) is displayed, 
perform the procedure in Sectio3.1 Selecting Internal or External Calibration and set the Cal 
Type to Internal
4.  Insert the probe tip into the CAL Port within 10 seconds 
after pressing the CAL button. Be sure that the probe tip 
makes contact with the spring switch. At this time the 
calibration process should automatically start.  
 
The top number of the Calibration screen shows the 
“target” value (leak source rate), while the bottom 
number shows the current measured value. Keep the 
probe tip firmly against the back of the port for the 
duration of the calibration process, during which time the 
instrument is calibrated at both a high and low flow rate.  
If the probe tip looses contact with the spring switch before the 
calibration process completes, the message “CAL ABORTED” is 
displayed. At this time press any key to retain the previous 
calibration curve and return to the Data Display screen. 
5.  After a successful calibration the new calibration data is saved 
and the message “CAL OK” followed by “WITHDRAW PROBE 
FROM LEAK SOURCE” will appear. 
6.  Withdrawing the probe causes the display to switch back to the 
Data Display screen. Note, however, if the PCHK Mode is set to 
AUTO (refer to Section 2.4.7 PCHK Mode), then a pressure check 
is performed prior to the display of the Data Display screen. 
 
NOTE:
 After being calibrated the instrument is automatically 
placed into its Search mode. 
 
CAL 
OK 
Oz/yr 
R134A
SETUP 
INT  CAL 
WITHDRAW PROBE
FROM LEAK 
SOURCE 
Oz/yr 
R134A
SETUP 
INT  CAL
Oz/yr 
R134A 
SETUP 
SM  SRCH
0.00 
Oz/yr 
R134A
SETUP 
INT  CAL
TARGET 
O/YR 
R134A 
MEASURED
O/YR
0.89 
Oz/yr 
R134A 
SETUP 
INT  CAL 
0.89 
TARGET 
O/YR 
R134A 
MEASURED 
O/YR 
INSERT 
PROBE 
 
 
 
PCHK