Bacharach H25-IR User Manual

Page of 60
H25-IR 
Maintenance 
 
Instruction 3015-4342 
41
 
5.2.6 Leak TEMPCO 
 
 
 
 
The Leak Temperature Coefficient value is set to match the temperature coefficient of the internal leak 
source as indicated on the tag attached to the leak source. Use the Left and Right buttons to move the 
blinking cursor across the screen to highlight the digit to be modified, and then use the Up and Down 
buttons to modify that digit. Press ENT to save the new Leak Temperature Coefficient value. 
TIP:  
To convert a temperature coefficient from %/degF to % degC, multiply the  
degF value by a factor of 1.8. For example: 0.3%/degF x 1.8 = 0.54%/degC. 
5.2.7 Sensor Cal 
 
 
 
 
The Sensor Calibration function allows the service technician to manually calibrate the instrument’s 
Pressure, Flow Rate, and Box Temperature sensors. Use the Up and Down buttons to highlight the 
sensor to be calibrated, and then press ENT to display that sensor’s calibration screen. 
5.2.7.1  Pressure Sensor Calibration Screen 
 
 
 
 
Calibrate the pressure sensor by first measuring the ambient atmospheric pressure using an appropriate 
pressure measurement device calibrated to read PSIA. Next, adjust the value displayed on the Pressure 
Sensor Calibration screen to match the measured pressure reading by using the Left and Right buttons 
to move the blinking cursor across the screen to highlight the digit to be modified, and then using the Up 
and Down buttons to modify that digit. Press ENT to save the new ambient pressure value. 
Tip:
 To convert inHg into PSIA, multiply inHg by 0.491. For example: 30.00 inHg x 0.491 = 14.73 PSIA. 
5.2.7.2  Flow Rate Sensor Calibration Screen (Factory Calibration) 
 
 
 
 
Set the default factory calibration flow and pressure drop values by first performing an internal 
calibration as described in Section 3.2 Internal Calibration
INT:
 Flow rate in moles/minute for both the Search and Measure modes after performing a calibration 
using the internal leak source, along with showing the manifold’s pressure drop.  
FAC:
 Flow rate and pressure drop values that are currently stored as the default factory calibration 
values. These values are used to validate calibrations performed by the operator.  A calibration error will 
occur if the operator’s calibration and factory values differ by more than 20%. 
Press ENT to store the internal calibration numbers as factory values. These are the calibration values 
that are reset when RESTORE FACTORY CAL is selected from the CAL Data menu (refer to Sectio3.5 
 
LEAK TEMPCO 
0
.54 
                           %/degC 
PRESSURE 
FLOW RATE 
BOX TEMP 
ENTER AMBIENT PSIA 
             1
4
.18 
      
CURRENT OFFSET=0.00 
S-Mpm/dP M-Mpm/dP 
INT 0.0177/-1.33 
0.0100/-2.28 
FAC 0.0177/-1.33 
0.0100/-2.28 
ENTER TO SAVE INT AS FACTORY