Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Basic Operation 
 
99 
6.6.3. REPORTING RANGE
 
MODES 
The M300E/EM provides three analog output range modes to choose from.   
 Single range (SNGL)  mode sets a single maximum range for the analog output.  If single range is selected 
both outputs are slaved together and will represent the same measurement span (e.g.  0-50 ppm), 
however their electronic signal levels may be configured for different ranges (e.g.  0-10 VDC vs.  0-.1 
VDC). 
  Dual range (DUAL) allows the A1 and A2 outputs to be configured with different measurement spans as 
well as separate electronic signal levels. 
  Auto range (AUTO) mode gives the analyzer to ability to output data via a low range and high range.  
When this mode is selected the analyzer will automatically switch between the two ranges dynamically as 
the concentration value fluctuates.   
Range status is also output via the external digital I/O status outputs (See Section 3.3.3). 
To select the Analog Output Range Type press: 
 
 
NOTE 
Upper span limit setting for the individual range modes are shared.  Resetting the span limit in one mode 
also resets the span limit for the corresponding range in the other modes as follows: 
 SNGL  DUAL   AUTO  
 Range 

 Range1   Low 
Range 
  
 
Range2 

 High 
Range   
 