Teledyne M300EM User Manual

Page of 398
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Basic Operation 
 
103 
6.6.3.3. RNGE  MODE  AUTO: Configuring the M300E/EM Analyzer for AUTO Range Mode  
In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges 
(low and high).   
  The unit will switch from low range to high range when the CO
2
 concentration exceeds 98% of the low 
range span.   
  The unit will return from high range back to low range once both the CO
2
 concentration falls below 75% of 
the low range span. 
In AUTO Range Mode the instrument reports the same data in the same range on both the A1 and A2 outputs 
and automatically switches both outputs between ranges as described above.   
Also the RANGE test function displayed on the front panel will be replaced by two separate functions: 
RANGE1: The LOW range setting for all analog outputs. 
RANGE2: The HIGH range setting for all analog outputs. 
The high/low range status is also reported through the external, digital status bits (See Section 3.3.3).  
To set individual ranges press the following keystroke sequence. 
SETUP X.X       
 CFG DAS
CLK MORE
EXIT
SETUP X.X     
SET   UNIT  DIL
EXIT
SETUP X.X     
SNGL DUAL
 DIL
ENTR EXIT
SETUP X.X     
 SNGL DUAL AUTO
EXIT
SETUP X.X     
 MODE
  UNIT  DIL
EXIT
SETUP X.X     
Toggle these keys to 
select the upper 
limit for the 
reporting range.
 discards the new 
setting.
 accepts the 
new setting.
SETUP X.X     
The
 and
ranges have separate 
slopes and offsets for 
computing the CO 
concentration. 
The two ranges must 
be independently 
calibrated
.
Avoid accidentally setting the
range (
) of the instrument 
with a higher span limit than the 
 range (
).  
This will cause the unit to stay in 
the low reporting range perpetually 
and defeat the function of the 
 range mode. 
 
04288D DCN5752