Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
224 
10.4.3. SYSTEM AUDIT/VALIDATION 
A system audit is an on-site inspection and review of the quality assurance activities used for the total 
measurement system (sample collection, sample analysis, data processing, etc.); it is an appraisal of system 
quality. 
Conduct a system audit at the startup of a new monitoring system and periodically (as appropriate) as significant 
changes in system operations occur. 
 
10.5. DYNAMIC MULTIPOINT CALIBRATION PROCEDURE 
10.5.1. LINEARITY TEST 
In order to record the instrument’s performance at a predetermined sensitivity and to derive a calibration 
relationship, a minimum of three reference points and one zero point uniformly spaced covering 0 to 90 percent 
of the operating range are recommended to define this relationship. 
The analyzer's recorded response is compared with the known concentration to derive the calibration 
relationship.  
To perform a precision check during the instrument set up, the sources of zero air and sample gas should conform to 
those described in Section 9.1.2. 
Follow the procedures described in Section 9.3 for calibrating the zero points.   
For each mid point: 