Theory of Operation 
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
230  
Lengthening the absorption path is accomplished partly by making the physical dimension of the reaction cell 
longer, but primarily by adding extra passes back and forth along the length of the chamber. 
Table 11-1:  Absorption Path Lengths for M300E and M300EM 
MODEL 
TOTAL NUMBER OF 
REFLECTIVE PASSES 
DISTANCE BETWEEN MIRRORS 
TOTAL 
ABSORPTION LIGHT 
PATH 
M300E 
32 
437.5 mm 
14 Meters 
M300EM 
312.5 mm 
2.5 Meters 
 
 
 
IR 
Source 
IR Beam 
Sample Chamber 
Band-Pass Filter 
Photo-Detector 
 
Figure 11-1: 
Measurement Fundamentals 
Upon exiting the sample cell, the beam shines through a band-pass filter that allows only light at a wavelength of 
4.7 µm to pass.  Finally, the beam strikes a solid-state photo-detector that converts the light signal into a 
modulated voltage signal representing the attenuated intensity of the beam.   
 
11.2.1. GAS FILTER CORRELATION 
Unfortunately, water vapor absorbs light at 4.7 µm too.  To overcome the interfering effects of water vapor the 
M300E/EM adds another component to the IR light path called a Gas Filter Correlation (GFC) Wheel. 
Measurement Cell
(Pure N
2
)
 
Reference Cell 
(N
2
 with CO)
 
 
Figure 11-2: 
GFC Wheel 