Teledyne M300EM User Manual

Page of 398
Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Troubleshooting & Repair 
 
293
 
13.5.6.6. Pressure/Flow Sensor Assembly 
The pressure/flow sensor PCA, located on the top of the absorption bench, can be checked with a voltmeter 
using the following procedure which, assumes that the wiring is intact, and that the motherboard and the power 
supplies are operating properly: 
1.  For Pressure related problems: 
  Measure the voltage across C1 it should be 5 ± 0.25 VDC.  
   If not then the board is bad. 
  Measure the voltage across TP4 and TP1.   
  With the sample pump disabled it should be 4500 mV ±250 mV.   
  With the pump energized it should be approximately 200 mV less.  If not, then S1, the pressure 
transducer is bad, the board is bad, or there is a pneumatic failure preventing the pressure 
transducer from sensing the absorption cell pressure properly. 
2.  For flow related problems: 
  Measure the voltage across TP2 and TP1 it should be 10 ±0.25 VDC.  
   If not then the board is bad. 
  Measure the voltage across TP3 and TP1.   
With proper flow (800 sccm at the sample inlet) this should be approximately 4.5V (this voltage 
will vary with altitude).   
  With flow stopped (sample inlet blocked) the voltage should be approximately 1V.   
  If the voltage is incorrect, the flow sensor is bad, the board is bad or there is a leak upstream of 
the sensor. 
04288D DCN5752