Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Getting Started 
 
47 
3.5.1. STARTUP 
After the electrical and pneumatic connections are made, turn on the instrument.  The pump and exhaust fan 
should start immediately.  The display should immediately display a single, horizontal dash in the upper left 
corner of the display.  This will last approximately 30 seconds while the CPU loads the operating system. 
Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data. 
During this process, astring of messages will appear on the analyzer’s front panel display. 
System waits 3 seconds then 
automatically begins its 
initialization routine.
.
System is checking the validity and 
functionality of the Terminus 
memory and firmware.
If at this point,
appears, contact Teledyne 
Instruments  customer service
.
The instrument is loading 
configuration and calibration data 
from  the flash chip.
The startup process may hesitate at this point if:
·
The Ethernet option is installed; 
·
DHCP mode is turned on and; 
·
The instrument is not connected to a 
functioning network.
The instrument is loading the 
system firmware.
 
The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and start 
monitoring CO gas. 