Teledyne API – Technical Manual - Model 300E Family CO Analyzers 
Optional Hardware and Software 
 
67 
5.6.3. ZERO/SPAN/SHUTOFF VALVE (OPTION 50B) 
This option requires that both zero air and span gas be supplied from external sources.   
  Span gas will be supplied from a pressurized bottle of calibrated CO gas.  
  A critical flow control orifice, internal to the instrument ensures that the proper flow rate is maintained.   
  An internal vent line ensures that the gas pressure of the span gas is reduced to ambient atmospheric 
pressure.   
  A  SHUTOFF valve preserves the span gas source when it is not in use. 
  Zero gas is supplied by either an external scrubber or a zero air generator such as the Teledyne API’s 
M701.   
5.6.3.1. Internal Pneumatics (OPT 50B) 
 
Figure 5-5: 
Internal Pneumatic Flow OPT 50B – Zero/Span/Shutoff Valves  
Table 5-2:  Zero/Span Valve Operating States for Option 50B 
MODE VALVE 
CONDITION 
Sample/Cal 
Open to SAMPLE inlet 
Zero/Span 
Open to IZS inlet 
SAMPLE 
(Normal 
State) 
Shutoff Valve 
Closed 
Sample/Cal 
Open to ZERO/SPAN valve 
Zero/Span 
Open to IZS inlet 
ZERO CAL 
Shutoff Valve 
Closed 
Sample/Cal 
Open to ZERO/SPAN valve 
Zero/Span 
Open to SHUTOFF valve 
SPAN CAL 
Shutoff Valve 
Open to PRESSURE SPAN Inlet 