Section 3: Configuration, Startup and Operation
6888Xi Advanced Electronics Instruction Manual
December 2012
PN 51-6888Xi
TX  IOB  Parameter Name
Unit
Description
N
Y
Heater Latch Off
--
Heater latch Off. (Parameter for Extended Temperature feature.)
No = The heater will turn back on once the temperature falls 
below the heater set point
Yes = The heater will not turn back on once the temperature falls 
below the heater set point
N
Y
Features
--
Advanced software features: 
0 = Stoichiometer, 1 = Prog Reference, 2 = Extended Temp, 4 = Diffuser Warning
Y
Y
O2 LRV
%
Primary variable (O2 %) lower range value.
Y
Y
O2 URV
%
Primary variable (O2 %) upper range value
N
Y
AO Range
--
Analog output polarity:
0 = 4-20mA, 1 = 20-4mA
Y
Y
Alarm Level
--
O2 alarm level:
0 = 3.5 mA, 1 = 21.1mA
N
Y
Relay 1
--
Alarm Relay 1 mode:
0 = Unit alarm, 1 = Low O2, 2 = Cal Recommended, 3 = In Calibration
N
Y
Relay 2
--
Alarm Relay 2 mode:
0 = Unit alarm, 1 = Low O2, 2 = Cal Recommended, 3 = In Calibration, 
4 = Heater Relay (Default)
N
Y
AO Track
--
Analog output track O2 sensor measurement during a calibration: No/Yes
N
Y
Tol Check
--
Enable/Disable tolerance check during calibration:  No/Yes
N
Y
Cal Recommend
--
Enable/Disable calibration recommended alarm:  No/Yes
N
Y
Cal Acknowledge
--
Enable/Disable calibration acknowledge before update calibration 
constants:  No/Yes
Y
Y
Cal Gas 1
%
Test Gas 1 value. This is the actual value of the gas being applied during the Test
Gas 1 phase of a calibration.
Y
Y
Cal Gas 2
%
Test Gas 2 value. This is the actual value of the gas being applied during the Test
Gas 2 phase of a calibration.
Y
Y
Gas Time
sec
Test Gas application time. This is the length of time test gases are applied to the
O2 probe during low or high Test Gas phase of a calibration.
Y
Y
Purge Time
sec
Test Gas purge time. This is the length of time before the output will be returned
to the process reading after a calibration.
N
Y
Plugged Diffuser
--
Enable Plugged Diffuser functionality: (Parameter for Diffuser Warning feature.)
0 = Disable, 1 = Rate, 2 = Proximity, 3 = Rate & Prox
N
Y
Auto Advance Cal
--
Auto advance calibration state machine when it detects that the cell response has
stabilized during test gas flow and purge: No/Yes
(Parameter for Diffuser Warning feature.)
N
Y
Enable Auto Cal
--
Enable/disable automatic calibration: No/Yes
N
Y
Start on Cal Rec
--
Start automatic calibration on Cal Recommended state: No/Yes
N
Y
Cal Interval
hr
Automatic calibration interval. The number 9999 disable the automatic 
timed calibration.
N
Y
Next Cal Time
hr
Time remaining until the next automatic periodic calibration. The number 9999
disable the next automatic timed calibration.
N
Y
Resource |
Analog Output
I/O board analog output resource mapping. One for each probe:
0x00000001 is for IOB AO
0x00000002 is for TX AO
N
Y
Resource | Auto Cal
I/O board auto calibration resource mapping. One for each probe:
0x00000100 is for SPS
0x00000200 is for IMPS
N
Y
Resource | Relay 1
I/O board’s relay 1 resource mapping. One for each probe:
0x00010000 is for IOB RELAY 1
N
Y
Resource | Relay 2
I/O board’s relay 2 resource mapping. One for each probe:
0x02000000 is for IOB RELAY 2
Table 3-5. Probe Parameters (3 of 4)