Emerson Process Management 760004-A User Manual
Instruction Manual
760004-A
February 2002
February 2002
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 HFID
• Power supply test - Verifies that all
internal DC voltages are within the
required tolerances.
required tolerances.
• Network test - Checks the internal
network interface.
• 20 bit ADC test - Checks the 20-bit
ADC on the Analysis Computer PCB
by sending a DC signal through the
Preamp PCB and reading the signal
back with the 20-bit ADC.
by sending a DC signal through the
Preamp PCB and reading the signal
back with the 20-bit ADC.
• 12 bit ADC test - Checks the 12-bit
ADC on the Analysis Computer PCB
by sending a DC signal and reading
the signal back with the 12-bit ADC.
by sending a DC signal and reading
the signal back with the 12-bit ADC.
• Power Supply PCB test - Checks
the presence of the Power Supply
PCB by activating the 3-way air sole-
noid.
PCB by activating the 3-way air sole-
noid.
• Safety PCB test - Checks the pres-
ence of the Safety PCB by sending a
command and reading it back.
command and reading it back.
• Case temperature test - Compares
the temperature read between the
Preamp temperature sensor and the
case temperature sensor. They must
be within 10
Preamp temperature sensor and the
case temperature sensor. They must
be within 10
°C of each other. This test
sometimes fails if the case is opened.
The sensor in the Preamp will take
longer to cool off since it is in an en-
closure. Re-running the self-test after
thermal equilibrium will produce a
positive result if the sensors are
working properly.
The sensor in the Preamp will take
longer to cool off since it is in an en-
closure. Re-running the self-test after
thermal equilibrium will produce a
positive result if the sensors are
working properly.
• Oven/Sample Temperature test -
Compares the temperature read be-
tween the sample temperature sensor
and the oven temperature sensor.
They must be within 50
tween the sample temperature sensor
and the oven temperature sensor.
They must be within 50
°C of each
other.
The self-test can be repeated at any time
by activating the TEST softkey in the Self
Test Results menu.
by activating the TEST softkey in the Self
Test Results menu.
9. Set the desired oven setpoint in the range
of 93
°C to 204°C (200°F to 400°F).
10. Wait for the Purge Air green LED to illu-
minate.
11. Introduce the remaining supply gases.
Perform leak check. (See Section 1-5
Specifications)
Specifications)
12. Set and verify the internal gas pressures.
Internal
Pressure
Regulator
Typical Operating
Pressures
Burner Air
965 to 1103 hPa-gauge
(14 to 16 psig)
(14 to 16 psig)
Fuel
1516 to 1723 hPa-gauge
(22 to 25 psig)
(22 to 25 psig)
Sample
(non-adjustable)
(non-adjustable)
206 to 290 hPa-gauge
(3.0 to 4.0 psig)
(3.0 to 4.0 psig)
Purge air of the following specifications
must be present:
must be present:
Flow: 16 to 18 L/min.
Supply Pressure: 689 to 1378 hPa-gauge
(10 to 20 psig)
Noncompliance could cause damage to
the module. At the very least, the mod-
ule's safety system, which requires a
certain volume of purge air flowing
through the case before allowing burner
ignition, will not allow the instrument to
operate. The lowest purge air
flow/pressure setting possible during
burner operation is preferable. Thus, the
user should set the external purge air
pressure initially at 689 hPa-gauge (10
psig). Check the Miscellaneous Control
Parameters screen under Technical Di-
agnostics, and note whether the Purge
Gas (switch) variable is "ON." If it is
"OFF," increase purge air supply by
69 hPa-gauge (1 psig), and recheck the
Purge Gas variable until it reads "ON."
DO NOT EXCEED 1378 hPa-GAUGE (20
PSIG). If the maximum setting is reached,
the module. At the very least, the mod-
ule's safety system, which requires a
certain volume of purge air flowing
through the case before allowing burner
ignition, will not allow the instrument to
operate. The lowest purge air
flow/pressure setting possible during
burner operation is preferable. Thus, the
user should set the external purge air
pressure initially at 689 hPa-gauge (10
psig). Check the Miscellaneous Control
Parameters screen under Technical Di-
agnostics, and note whether the Purge
Gas (switch) variable is "ON." If it is
"OFF," increase purge air supply by
69 hPa-gauge (1 psig), and recheck the
Purge Gas variable until it reads "ON."
DO NOT EXCEED 1378 hPa-GAUGE (20
PSIG). If the maximum setting is reached,