Emerson Process Management 3422 User Manual

Page of 398
 Reference, Installation, and Operations Manual
Section 5: Startup and meter configuration
3-9000-743 Rev S
June 2013
Startup page
 135
Pressure expansion correction: Only enabled if Live or Fixed was selected for 
Pressure for meter corrections and Pressure expansion correction was selected 
on the Startup Page.
Pipe outside diameter: Enter the outside diameter of the meter body.
Youngs modulus: Enter the Young's Modulus value (ratio of tensile stress to 
tensile strain)
Poissons ratio: Enter the Poisson's Ratio value (the absolute ratio of the pipe 
material lateral strain over axial strain) 
Table 5-12  Enabling Meter Corrections
Daniel MeterLink
 
Display Name
Data Points, Options and Guidelines
Base condition
correction
Data points affected:
None (enables/disabled access to other Daniel MeterLink configuration screens, data 
fields)
Options:
Disabled if unchecked
Enabled if checked
Guidelines:
This selection enables/disables access to other Daniel MeterLink configuration screens 
and/or data fields.
For options and guidelines, see Daniel MeterLink online help.
Flow profile correction 
(for JuniorSonic
TM 
only)
Data points affected:
None (enables/disables access to other Daniel MeterLink configuration screens, data 
fields)
Options:
Disabled if unchecked
Enabled if checked
Guidelines:
For options and guidelines, see Daniel MeterLink online help.
Temperature expansion correc-
tion
Data points affected:
EnableExpCorrTemp
Options:
Disable (FALSE)
Enable (TRUE)
Guidelines:
For options and guidelines, see Daniel MeterLink online help.
For Daniel Mark II Gas Ultrasonic meters only, the Pressure expansion correction is only 
available for DFI boards running version 3.10 or later firmware.