Set calibration parameters
Section 5: Startup and meter configuration
 Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
Piece-Wise linearization- wet calibration
When the Piece-Wise linearization wet calibration method is selected, the meter uses a 12-point 
piece-wise linearization on the dry-calibration velocity. Up to twelve flow rate and meter factor 
pairs can be specified for each flow direction from the Daniel MeterLink 
Calibration - Meter 
Factors
 page as shown in the table below.
Table 5-24  Data Points for Piece-Wise 
linearization
 calibration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
Forward Flow Rate
Data points affected:
FwdFlwRate1…FwdFltRate12
Options:
Enter a value (m
3
/h or ft
3
/h) within the range [0, 200000 m
3
/h].
Guidelines:
The flow rates must be entered from highest to lowest.
For additional guidelines, se
 and Daniel MeterLink online help.
Forward Meter Factor
Data points affected:
FwdMtrFctr1…FwdMtrFctr12
Options:
Enter a value (dimensionless) within the range [0.95,1.05].
Guidelines:
For additional guidelines, se
 and Daniel MeterLink online help.
Reverse Flow Rate
Data points affected:
RevFlwRate1…RevFltRate12
Options:
Enter a value (m
3
/h or ft
3
/h) within the range [0, 200000 m
3
/h].
Guidelines:
The flow rates must be entered from highest to lowest.
For additional guidelines, se
 and Daniel MeterLink online help.
Reverse Meter Factor
Data points affected:
RevMtrFctr1…RevMtrFctr12
Options:
Enter a value (dimensionless) within the range [0.95,1.05].
Guidelines:
For additional guidelines, se
 and Daniel MeterLink online help.