2.3 Freeze-Pump-Thaw cycle 
 
Using this system it is possible to operate freeze-pump-thaw cycles on sample liquids 
held in any of the three vessels.  Procedures appropriate to each vessel are outlined 
below. 
 
2.3.1 Sample in Vessel 1 
 
•  All taps and valves should be closed.  
•  Open valve V2 and tap A on Vessel 1, and freeze the sample liquid by slowly 
raising a dewar of liquid nitrogen around it. 
•  When it has been determined that the liquid has completely frozen switch the 
pump to roughing mode and open valve V1. 
•  Switch the pump to turbo mode and leave for several minutes.  The pressure 
within the system can be read from the baratron gauge on the pump. 
•  When sufficient pumping has been done and the sample is ready to thaw, close 
valve V1 and remove the dewar of liquid nitrogen from around Vessel 1. 
•  The system must be ‘let-up’ to atmospheric pressure using helium gas by 
opening valves V3 and V9.  The pressure of gas in the system can be read 
from the baratron gauge. 
•  Tap B should be closed to prevent suck-back of the liquid as the solid thaws. 
•  The sample can be warmed gently to aid the thawing process and when 
complete close tap A. 
 
2.3.2 Sample in Vessel 2 
 
•  All taps and valves should be closed.  
•  Open valve V2 and tap E on Vessel 2 and freeze the sample liquid by slowly 
raising a dewar of liquid nitrogen around it. 
•  When it has been determined that the liquid has completely frozen switch the 
pump to roughing mode and open valve V1. 
•  Switch the pump to turbo mode and leave for several minutes.  The pressure 
within the system can be read from the baratron gauge on the pump. 
•  When sufficient pumping has been done and the sample is ready to thaw, close 
valve V1 and remove the dewar of liquid nitrogen from around Vessel 2. 
•  The system must be ‘let-up’ to atmospheric pressure using helium gas by 
opening valves V3 and V9.  The pressure of gas in the system can be read 
from the baratron gauge. 
•  Valve V7 should be closed to prevent suck-back of the liquid as the solid 
thaws. 
•  The sample can be warmed gently to aid the thawing process and when 
complete close tap E. 
 
 
 
 -