Bacharach AGM300 User Manual

Page of 94
64
 
Instruction 3015-4275 
AGM300 / ADM800 Maintenance 
Warnings and Cautions 
WARNING - Always remove AC input power to the monitor 
before working on the interior of the instrument.  
CAUTION - When working inside the instrument be very 
careful not to dislodge any electrical wiring or pneumatic 
tubing. The AGM300 contains sensitive electronic 
components that can be easily damaged. Be careful not to 
touch or disturb any of these components. 
 
Servicing Inline Filter  
Locate the inline filter (Page 6) and inspect it for obvious signs of contamination. If the filter is contaminated or 
has started to turn red in color it should be replaced with P/N 3015-3071. 
To replace the filter, remove it from the retaining clip and carefully slide the two pieces of tubing off the fittings. 
Re-fit the tubing on the new filter, making sure the flow arrow points to the right. Complete the installation by 
pushing the new filter back into the retaining clip. 
Servicing Air Lines 
System air lines should be checked periodically for obvious signs of kinks, damage, and contamination. 
Replace the tubing as required with P/N 304-2742 or equivalent. 
The end-line filters should be checked periodically and replaced when there are obvious signs of contamination. 
To service the filter simply remove it from the air line and replace it with P/N 3015-3420. 
Replacing Fuses 
Locate the fuse holders (Pages 7 & 23) and remove the fuses from their holders. Visually inspect and test the 
fuses with an ohm meter for continuity. If necessary, replace any defective fuses. 
AGM300, 120 VAC Input, Power Supply P/N 3015-3167 
F1: 0.5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2538 
F2: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537 
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539 
F4: 63 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2535 
 
AGM300, 230 VAC Input, Power Supply P/N 3015-3161 
F1: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537 
F2: 125 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2536 
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539 
F4: 32 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2534 
 
ADM800 
F1 & F2: 1 A, 250VAC, Fast Acting, 5mm x 20mm, P/N 3015-4011