Bacharach AGM300 User Manual

Page of 94
70
 
Instruction 3015-4275 
Summary of Registers 
 
Register 
Name 
Number Type 
Description 
     
System Data 
 
0x0010h 
R/W 
AGM System Setup Data 
Status 
  0x0011h R/W 
Operating summary of faults, alarms 
      and 
status 
Zone Data 
 
0x12xxh 
R/W 
Setup data for up to 16 zones (xx  
      defines 
zone 
number) 
CAL Data 
 
0x0014h 
R/W 
Cal Factors for all gases 
Date/Time 
 
0x0015h 
R/W 
Set AGM300 date & time 
Sensor Data 
 
0x0016h 
Raw measurement of sensors 
Rel. Hold 
 
0x0017h 
Release AGM300 out of hold mode 
Hold Zone 
 
0x0018h 
Put AGM300 into hold mode 
Fault Log 
 
0x19xxh 
20 most recent fault events  (xx = 00  
 
 
 
 
 
 
or 01) 
Alarm Log 
 
0x1Axxh 
20 most recent alarm events (xx = 00,  
      01, 
or 
02) 
Serv. Mode 
 
0x001Bh 
Puts AGM300 into service mode 
Rel Serv. 
 
0x001Ch 
Release AGM300 from service mode 
PPM    
 
0x001Eh 
PPM values for all zones 
Zone Log 
 
0x3yxxh 
Trend data for each AGM300 zone (y  
= zone number (starting at 0), xx = 00 – 06)Data  
 
Data Type Abbreviations 
 
C ...................... Character 
Float ................. Floating Point 
I ........................ Integer 
TIM................... Time 
UC.................... Unsigned Character 
UI ..................... Unsigned Integer