Agilent Technologies E4350B User Manual
80 Language Dictionary
[SOUR:]VOLT:TABL:OFFS
This command adds a voltage offset when operating in table mode.
Command Syntax
[SOURce]:VOLTage:TABLe:OFFSet <NRf>
Parameter
0 to VMAX
*RST Value
0
Examples
VOLT:TABL:OFFSet 4
Query Syntax
VOLT:TABL:OFFS?
Returned Parameters
<NRI>
Status Subsystem
This subsystem programs the Agilent SAS status registers. The Agilent SAS has three groups of status registers;
Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as
described in "Chapter 4 - Status Reporting". The Operation and Questionable status groups each consist of the Condition,
Enable, and Event registers and the NTR and PTR filters. See chapter 4 for more information about these registers.
Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as
described in "Chapter 4 - Status Reporting". The Operation and Questionable status groups each consist of the Condition,
Enable, and Event registers and the NTR and PTR filters. See chapter 4 for more information about these registers.
Status Operation Registers
Bit Configuration of Operation Registers
Bit Position
15-12
11
10
9
8
7
6
5
4
3
2
1
0
Bit Name
NU
NU
CC
NU
CV
NU
NU
WTG
NU
NU
NU
NU
CAL
Bit Weight
2048
1024
512
256
128
64
32
16
8
4
2
1
CAL = Interface is computing new calibration constants; CC = The power module is in constant current mode.
CV = The power module is in constant voltage mode; NU = (Not used); WTG = Interface is waiting for a trigger.
STAT:OPER?
This query returns the value of the Operation Event register. The Event register is a read-only register which holds (latches)
all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.
all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.
Query Syntax
STATus:OPERtion[:EVENt]?
Parameters
(None)
Returned Parameters
<NR1>
(Register Value)
Examples
STAT:OPER?
STATUS:OPERATIOBAL:EVENT?
Related Commands
*CLS STAT:OPER:NTR STAT:OPER:PTR
STAT:OPER:COND?
This query returns the value of the Operation Condition register. That is a read-only register which holds the real-time
(unlatched) operational status of the Agilent SAS.
(unlatched) operational status of the Agilent SAS.
Query Syntax
STATus:OPERation:CONDition?
Parameters
(None)
Examples
STAT:OPER:COND? STATUS:OPERATION:CONDITION?
Returned Parameters
<NR1> (Register value)
Related Commands
(None)