Agilent Technologies E4350B User Manual

Page of 129
Language Dictionary    85
TRIG
When the trigger subsystem is enabled, TRIG generates a trigger signal.  The trigger will then:
1.
Initiate a pending level change as specified by CURR[:LEV]:TRIG or VOLT[:LEV]:TRIG.
2.
Clear the WTG bit in the Status Operation Condition register.
3.
If INIT:CONT has been given, the trigger subsystem is immediately re-enabled for subsequent triggers.  As soon as it
is cleared, the WTG bit is again set to 1.
Command Syntax
TRIGger[:IMMediate]
Parameters
(None)
Examples
TRIG     TRIGGER:IMMEDIATE
Related Commands
ABOR    CURR:TRIG     INIT     *TRG    VOLT:TRIG
TRIG:SOUR
This command selects the trigger source.  Since the Agilent SAS has no other trigger source than the GPIB bus, this
command need not be used.  It is included in the command set to provide programming compatibility with other instruments
(such as the Agilent Electronic Load family) that may have more than one trigger source.
Command Syntax
TRIGer:SOURce <CRD>
Parameters
BUS
*RST Value
BUS
Examples
TRIG:SOUR  BUS      TRIGGER:SOURCE BUS
Query Syntax
TRIGger:SOURce?
Returned Parameters
BUS
Related Commands
*RST *TRG TRIG[:IMM]