 Model 2400S Transmitters 
I/O Wiring – Model 2400S Analog Transmitters
Hazardous area inductance
The inductance (Li) of the Model 2400S transmitter is 0.0 mH. This value plus the field wiring 
inductance (Lcable) must be lower than the maximum allowable inductance (L
a
) specified by the IS 
barrier. The following equation can then be used to calculate the maximum cable length between the 
transmitter and the barrier: 
3.5.2
Hazardous area mA wiring 
Figure 3-15 illustrates the basic hazardous area wiring for the Model 2400S Analog transmitter’s mA 
output.
Figure 3-15
Hazardous area mA output wiring 
L
i
L
cable
L
a
+
Hazardous area
Safe area
R
barrier
4–20 mA
R
load
V
in
V
out
Ground
Note: R
barrier
 and R
load
 should be added together to 
determine the proper V
in
R
barrier
Note: Channel B (terminals 3 and 4) can also be 
configured this way.