Fujimoto FKK 4 User Manual

Page of 74
64
(1) Make wiring according to the diagram below.
Connect DC power supply (power source), digital voltmeter (measuring device), standard
resistance and HHC (Hand Held Communicator):
When current is measured with an ammeter connected to CK+ and CK- terminals, the internal
resistance of the ammeter should be 12
Ω or less.
(2) Calibration of output circuit (D/A)
Calibrate the output circuit with reference to “Calibration of output circuit” (page 20) in “AD-
JUSTMENT ”.
Load resistance
Standard resistance
DC power supply
Digital voltmeter
HHC
Important
For communication with the HHC, a load resistor of 250
Ω is necessary.
Calibration procedure
(3) Zero/span adjustment
Refer to “Zero/span adjustment” (page 35) in Section 7.
(4) Accuracy test
Apply input pressures in the order of 0%, 25%, 50%, 75%, 100%, 75%, 50%, 25% and 0%,
and read output at each input pressure.
Make sure the difference between each output value and input pressure (%) is within the
accuracy rating listed in the table below.
 The voltage values in the table are dependent on use of “DC power supply + standard resistor
250
Ω + digital voltmeter (measuring device).
Measurement category
Reference value
Accuracy
Accuracy:0.1% Accuracy:0.2%
Percent display (%)
0, 25, 50, 75, 100
±0.1
 
±0.2
Current measurement (mA)
4, 8, 12, 16, 20
±0.016
±0.032
Voltage measurement (V)
1, 2, 3, 4, 5
±0.004
±0.008