A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
26 
 
8.1.1.
8.1.1.
8.1.1.
8.1.1.  ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode
ACCURACY of the Standard Mode and Quick Mode    
 
  ACCURACY of measurement can be set either 
 HI 
MID.
 or 
 LO 
 
  The sample quantity is automatically selected by ACCURACY. 
 
  The termination value of the analyzing mode is automatically selected by ACCURACY 
and minimum scale value of % display. 
 
  The analyzing mode is the program to finish the measurement when a change of 
moisture content per one minute becomes smaller than a preset termination value. 
 
  The settings are as follows: Specify an ACCURACY. 
 
  Drying rate: Change of moisture content per one minute [%/min] 
 
Standard Mode
Standard Mode
Standard Mode
Standard Mode    
    
ACCURACY 
 Model 
Minimum 
scale
HI 
MID. 
LO 
  0.001 % 
 0.01 %/min
 0.02 %/min   0.05 %/min
  0.01 % 
 0.02 %/min
 0.05 %/min   0.10 %/min
MS-70 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
  0.01 % 
 0.02 %/min
 0.05 %/min   0.10 %/min
MX-50 
  0.1 % 
 0.10 %/min
 0.20 %/min 
  0.05 % 
 0.05 %/min
 0.10 %/min 
 0.1 
MF-50 
 1 
 0.1 
Preset 
Termination 
value 
ML-50 
 1 
 0.10 %/min
 0.20 %/min 
 0.50 %/min
Sample quantity 
10 g 
5 g 
1 g 
Use Precise 
result 
 Quick 
measurement
 
 
Quick Mode
Quick Mode
Quick Mode
Quick Mode    
    
ACCURACY 
 Model 
Minimum 
scale
HI 
MID. 
LO 
  0.001 % 
 0.02 %/min
 0.05 %/min   0.05 %/min
  0.01 % 
 0.05 %/min
 0.10 %/min   0.20 %/min
MS-70 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
  0.01 % 
 0.05 %/min
 0.10 %/min   0.20 %/min
MX-50 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
 0.05 
 0.1 
MF-50 
 1 
 0.10 %/min
 0.20 %/min   0.50 %/min
 0.1 
Preset 
Termination 
value 
ML-50 
 1 
 0.20 %/min
 0.50 %/min   1.00 %/min
Sample quantity 
5 g 
2 g 
1 g 
Use Precise 
result 
 Quick 
measurement