A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
41 
 
9.
9.
9.
9.  Check Function
Check Function
Check Function
Check Function    
 
9.1.
9.1.
9.1.
9.1.  Self
Self
Self
Self----Check Function 
Check Function 
Check Function 
Check Function      
 
 
  Use the self-check function to check whether there is an error or inaccurate result. 
During the check, the heater is turned on and the temperature sensor is checked. 
 
Caution 
 
 
Do not put flammable matter near the analyzer. 
 
 
Do not put anything on the heater cover. 
 
 
9.1.1.
9.1.1.
9.1.1.
9.1.1.  Operation
Operation
Operation
Operation    
 
 
1.  Display the gram unit (of the weighing mode). 
 
 
 
2.  Press and hold the 
 PROGRAM 
 key to 
display   CH . 
 
 
  3.  Put the breeze break ring, pan support, pan 
handle and sample pan in order. (Do not put 
a sample on the pan.) 
Close the heater cover. 
Press the 
 ENTER 
  to start the check. 
 
  
If 
 Clo5e 
 is displayed, the heater cover is 
not closed. When it is closed, the self-check 
function is started. 
 
 
 
 
4.  The check function needs approximately one 
minute. 
 
  
Good 
result ...Displays 
 CH pa55 
, sounds 
buzzer and returns to weighing 
mode automatically. 
 
 
  Error..............The buzzer sounds and an error 
code is displayed.   
 
   
Refer to 15.5. Error Message for details. 
 Example: 
 CH no   
 
 
   
 error 0 
 
 
   
 Ht err  
 
 
 
 0.000
g
 
 CH
 
  CH . . . .
 
 0.000
g
 
Preparation 
 CH 
pa55
 
Checking 
Do not put sample 
Press and hold 
Good result 
Weighing mode