A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
50 
 
 
11.1.
11.1.
11.1.
11.1. RS
RS
RS
RS----232C Serial Interface
232C Serial Interface
232C Serial Interface
232C Serial Interface    
 
RS
RS
RS
RS----232C Serial Interface
232C Serial Interface
232C Serial Interface
232C Serial Interface    
 
  Transmission system 
EIA RS-232C 
 
  Transmission form 
Asynchronous, bi-directional, half duplex 
 
  Data format 
Baud rate 
2400bps 
  
Data 
bits 
7bits 
  
Parity 
EVEN 
  
Stop 
bit 
1bit 
  
Code 
ASCII 
 
  Terminator 
CR LF (CR: 0Dh, LF: 0Ah) 
  
Bit 
format 
 
 
 
 
 
 
 
 
 
Pin Connections
Pin Connections
Pin Connections
Pin Connections    
 
 
2.6mm, metric screw thread 
 
 
 
 
 
 
 
 
 
MX-50 and MF-50 (DCE) 
Computer (DTE) 
Pin No. 
Signal Name  2
Description 
Direction
Signal Name 
1 FG 
Frame 
ground 
FG 
2 RXD 
Receive 
data 
 TXD 
3 TXD 
Transmit 
data 
 RXD 
RTS 
Ready to send  3 
 RTS 
CTS 
Clear to send    3 
 CTS 
DSR 
Data set ready 
 DSR 
7 SG 
Signal 
ground 
- SG 
16, 18, 19, 
21, 23 
Internal use 
Do not connect    1 
Other Not 
used 
 
 
 
1:  Normal DOS/V cables do not use these terminals. 
 
 
2: Signal names of the analyzer side are the same as the DTE side with TXD and 
RXD reversed. 
 
 
3:  RTS and CTS control are not used. CTS output is HI always. 
 
 
RS-232C
-5 V to -15 V 
+5 V to +15 V 
Data bits 
Start bit 
Parity bit 
Stop bit