A&D MS-70 User Manual

Page of 75
MS-70 / MX-50 / MF-50 / ML-50 
59 
 
 
 
  10.  After fifteen minutes, the buzzer sounds and the display 
blinks 
  "1"6"0°C 
. Adjust the blinking value to the 
thermometer value using the 
 or 
 key. (Example: 162°C) 
 
 
 11. Press the 
 ENTER 
  key to store the new data, to finish 
the adjustment and to return to the weighing mode. 
 
 
  If the report for GLP, GMP and ISO is output,   
 glp 
  is displayed. Output condition is 
selected in the function table. 
 
 
 
 
 
 
 
 
 
 
  When the heater cover is opened during measurement or 
the 
 STOP 
  key is pressed, calibration is stopped and the 
analyzer displays the weighing mode. 
 
 
 
Calibration R
Calibration R
Calibration R
Calibration Report
eport
eport
eport Example for Temperatur
 Example for Temperatur
 Example for Temperatur
 Example for Temperature Sensor Adapted to GLP, GMP and ISO
e Sensor Adapted to GLP, GMP and ISO
e Sensor Adapted to GLP, GMP and ISO
e Sensor Adapted to GLP, GMP and ISO    
Preset the following parameters to print data to AD-8121B 
Device Parameter 
Description 
Interval 
pU5e 1  1 Approx. two seconds interval in each line. 
Analyzer 
Output format 
adapted to GLP, 
GMP and ISO 
info 1 
or 
info 2 
Calibration report is printed at "Calibration of the 
Weighting Sensor" and "Calibration of Drying 
Temperature". 
AD-8121B 
MODE 3 
Dump print (Received data is printed as it is) 
 
1: Factory 
settings 
 
 
...........
A & D ........
Manufacture 
 MODEL
......
MX-50 ........
Model 
 S/N
.....
K1234567 ........
Serial number 
 ID
.......
LAB-123 ........
ID number 
 DATE
..
2004/09/30 ........
Date 
 TIME
....
12:34:56 ........
Time 
 CALIBRATED 
 
 
.....
TEMPERATURE 
 
 TARGET
....
ACTUAL 
 
 100
..
C
.....
97
..
C ........
100°C target value 
measurement value 
 160
..
C
....
162
..
C ........
160°C target value 
measurement value 
 SIGNATURE 
 
 
 
........
Signature 
 
 
 
- - - - - - - - 
 
 
Buzzer
160℃  "1"6"0℃
 
160℃  "1"6"2℃
 
 end
 
 glp
 
 0.000
g
 
 0.000
g
 
Calibration type