Fluke 566 User Manual

Page of 30
566/568 
Users Manual 
The reading includes: 
  IR temperature 
  Thermocouple temperature (if connected) 
  Emissivity  
  Min/Max/Avg/Dif (if either Min/Max or Avg/Dif is enabled) 
  Date/Time 
You may also press the Cancel softkey to abort saving the reading. 
Light 
The thermometers are equipped with a backlit display with two brightness levels. The 
Light softkey is used to adjust the backlight brightness. The backlight is on each time the 
trigger is pulled.  
To toggle the backlight brightness, press the Light softkey. 
The backlight can be disabled using the Setup menu. See “Setup” for more information. 
Memory  
The thermometers can store measurement records including time, date, emissivity, and 
measurement record numbers (see “Save” for more information). The 566 can store 20 
records and the 568 can store 99. 
To access records stored in memory: 
1.  Press the Menu softkey until Mem appears as the left softkey function. 
2.  Press the Mem softkey to access the Memory menu. Saved readings can be read. 
Emissivity Menu 
The Emissivity menu includes a list of pre-defined materials and lists their typical 
emissivity values, see Table 3. For further information, see “Emissivity”. 
Note 
Default emissivity is 0.95. 
To access the Emissivity menu: 
1.  Press the Menu softkey until I (emissivity) appears as the right softkey function. 
2.  Press the I softkey. 
The Emissivity list may be accessed by pressing the Table softkey or a material’s typical 
emissivity may be entered manually by pressing the No. softkey. 
  If the Emissivity table is accessed, a listing of materials and their suggested 
emissivity is shown.  
1.  Use the down arrow to navigate through the list.  
2.  Press the Enter softkey to choose the desired material.