Fluke 566 User Manual

Page of 30
566/568 
Users Manual 
20 
Specifications 
Feature 566 
568 
IR Temperature Range 
-40 
°
C to 650 
°
(-40 
°
F to 1200 
°
F) 
-40 
°
C to 800 
°
(-40 
°
F to 1470 
°
F) 
Accuracy  
<0 °C: ±(1.0 °C + 0.1 °/1 °C) 
>0 °C: ±1 % or ± 1.0 °C, whichever is greater  
(<32  F±2 °F ±0.1 °/1 °F) 
>32 °F: ±1 % or ±2 °F), whichever is greater 
Repeatability 
±
0.5 % of reading or 
±
0.5 °C (1 °F), whichever is greater. 
Display Resolution 
0.1 
°
C / 0.1 
°
Spectral Response 
8 to 14 
μ
Response Time (95%) 
<500 ms 
K-Type Thermocouple Input 
Temperature Range 
-270 
°
C to 1372 
°
C (-454 
°
F to 2501 
°
F) 
K-Type Thermocouple Input 
Accuracy 
-270 
°
C to -40 
°
C: 
±
(1 °C + 0.2 °/1 °C) 
(-454 
°
F to -40 
°
F: 
±
(2 °F + 0.2 °/1 °F)) 
-40 
°
C to 1372 
°
C: 
±
1 % or 1 °C (-40 
°
F to 2501 
°
F: 
±
1 % or 2 °F), 
whichever is greater 
K-Type Thermocouple 
Resolution 
0.1 
°
C/0.1 
°
Distance:Spot  
(90% energy) 
30:1 50:1 
Laser sighting 
single laser, output <1 mW Class II, wavelength  
630 to 670 nm 
Minimum spot size 
19 mm 
Emissivity 
Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of 
common materials 
Data storage 
20 points 
99 points 
Communication 
none USB 
2.0 
Operating Altitude 
3000 meters above mean sea level 
Storage Altitude 
12,000 meters above mean sea level 
Relative Humidity 
10 % to 90 % RH non-condensing up to 30 °C (86 °F) 
Operating Temperature 
0 °C to 50 °C (32 °F to 122 °F) 
Storage Temperature 
-20 °C to 60 °C (-4 °F to 140 °F) 
Vibration 
2.5 G, IEC 68-2-6 
Weight 
0.322 kg (0.7099 lb) 
Dimensions 
17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W 
Power 
2 AA /LR6 Batteries 
2 AA /LR6 Batteries 
or USB connection when used 
with a PC