6-16  Calibration
ChartScan User’s Manual
 
2.  Apply calibration voltage from the precision calibrator to the first channel of each CSN14/HV/S card
(see following table), and verify with a digital multimeter.  Requirements for both devices are specified
in Required Equipment.
Volts DC
Slot # 1
Slot #  2
Slot #  3
Slot #  4
Slot # 5
Slot #  6
Slot #  7
Slot # 8
-2.5  V
G1,20X
G17,20 X
G33,20X
G49,20X
G65,20X
G81,20X
G97,20X
G113,20X
+2.5  V
G1,20X
G17,20X
G33,20X
G49,20X
G65,20X
G81,20X
G97,20X
G113,20X
-25.0  V
G1,21X
G17,21X
G33,21X
G49,21X
G65,21X
G81,21X
G97,21X
G113,21X
+25.0  V
G1,21X
G17,21X
G33,21X
G49,21X
G65,21X
G81,21X
G97,21X
G113,21X
-250.0  V
G1,22X
G17,22X
G33,22X
G49,22X
G65,22X
G81,22X
G97,22X
G113,22X
+250.0  V
G1,22X
G17,22X
G33,22X
G49,22X
G65,22X
G81,22X
G97,22X
G113,22X
Note:  In the slot columns, the digit after the G is the channel number and the number immediately preceding
the X assigns the voltage value, e.g., G1,20X means “Calibrate Gain for channel 1, ±2.5 V.”
3.  After each calibration type 
U2X
 and check for a return of  
U128
 to confirm no errors.
Response codes and meanings applicable to calibration are as follows:
Response to 
U2X
Indication for Calibration Mode
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature sensor out of range)
U144
Invalid password
U160
Invalid command
 
4.  After completing the gain calibration of all cards, type 
EX
 to end the calibration mode.
5.  Type 
U2X
 to confirm calibration mode is disabled.  A return of 
U000
 confirms this.
 Note: You can view the calibration constants by typing an entry similar to the following:
 
C#1X
  
QC?    
…for slot #1
 You can monitor the calibration voltage by typing an entry similar to the following:
 
*B        
to clear the buffer
 
C1,1X     
to addresses the channel, this example is for slot #1, channel 1
 
T1,1,0,0X;@X 
… to trigger the acquisition (Trigger LED flashes); ); 
@X
 starts acquisition
 
U13X 
every entry of U13X results in an updated value of voltage in counts
 Note that 
@X 
can be used to start or stop the acquisition.