Electrical Connections 
10 
Installation and Maintenance Manual 
10/07 
 
 
4. Electrical Connections 
4.1 Overview 
The terminal board connections for the various Honeywell measuring instruments are given in the 
appropriate Figures in this section.  
To avoid the possibility of AC pickup in the cell leads, separate them from all AC line-voltage wiring or 
run them in a separate grounded conduit.  
ATTENTION 
Do not use shielded cable except where shown in the following figures. 
 
 
4.2  Instrument Wiring for 4973 Cells with Integral Cable 
4.2.1  Model 4973 Series with Integral Cable to UDA2182 Analyzer 
 
 
 
Figure 4-1 Installation Diagram, 4973 Cells, with junction box head connected to UDA2182 
Analyzer 
NOTES 
1. 
FOR PURE WATER SAMPLES IN NON-CONDUCTIVE (PLASTIC, GLASS, ETC.)  PIPING, GROUND 
THE BLACK CELL ELECTRODE LEAD NEAR THE CELL. ALTERNATIVELY, CONNECT TO THE UDA 
GROUND SCREW AS SHOWN DOTTED. DO NOT GROUND 10, 25, OR 50 CONSTANT CELLS. 
2. 
FOR CELL LEADS BLACK AND WHITE, USE 16 TO 22 AWG CABLE, SHIELDED TWISTED PAIR, 
WITH 30 pF MAX. CAPACITANCE BETWEEN CONDUCTORS. CONNECT SHIELD TOTERMINAL 
“10”. FOR MODEL 4973 INSTALLATIONS WITH HONEYWELL 31079198 SS FLOW CHAMBER, USE 
22 AWG MINIMUM COAXIAL CABLE TYPE RG59U (BELDEN 9259 OR EQUIV.). CONNECT COAX 
SHIELD TO TERMINALS “A” AND “10”. CONNECT COAX CONDUCTOR TO TERMINALS “C” AND “7”. 
3. 
FOR COMPENSATOR LEADS B AND D, USE 16 TO 22 AWG, TWO CONDUCTOR CABLE. 
4. 
CELL TO ANALYZER CABLES ARE CONSIDERED LOW LEVEL. RUN SEPARATE FROM HIGH 
LEVEL WIRING. 
DRAWING 50012874