Anritsu S113C User Manual

Page of 95
Out-of-Band Spurious Emission Measurements
Out-of-band spurious measurements are made on signals outside the system main band.
These signals, which can interfere with other communication systems, can be categorized
into harmonics and random spurious emissions. Real time monitoring of spurious emissions
from a transmitter can uncover unwanted signals before they interfere with other users on
the other channels.
Required Equipment
q
Site Master Model S114C or S332C
q
30 dB, 50 Watt, bi-directional, DC –18 GHz, N(m) – N(f), Attenuator, 42N50A-30
q
Test Port extension cable, Anritsu 15NNF50 – 1.5C
Procedure
Step 1. Using the test port extension cable and 30 dB attenuator connect the Site Master
to appropriate transmit test port.
Step 2. Press the
ON/OFF
key on the Site Master.
Step 3. Press the
MODE
key.
Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press
ENTER
.
Step 5. Select the center frequency for the channel of interest.
Step 6. Set the frequency span wide enough to include the primary channel and upper
and lower channel.
Step 7. Set the following parameters:
reference level = 0 dBm
input attenuation level = auto
resolution bandwidth =10 kHz
video bandwidth =300 Hz.
Step 8. Use the limit line and markers to read and compare the observed signals to the
specified allowable level of out-of-band spurious emissions for the correspond-
ing channel transmitted frequency.
5-6
Chapter 5 Spectrum Analyzer Measurements