Anritsu S113C User Manual

Page of 95
Chapter 6
Power Measurement
Introduction
The Site Master with Option 5 installed can be used for making power measurements with a
broadband (1 MHz to 3000 MHz) RF detector, Anritsu P/N 5400-71N50. The power moni-
tor displays the measured power results in dBm or Watts.
Power Measurement
Required Equipment
q
Site Master Model S114C or S332C with Power Monitor option
q
Broadband RF detector, Anritsu P/N 5400-71N50
q
30 dB, 50 Watt, bi-directional, DC –18 GHz, N(m) – N(f), Attenuator, 42N50A-30
Procedure
Step 1. Press the
ON/OFF
key on the Site Master.
Step 2. Press the
MODE
key.
Step 3. Use the Up/Down arrow key to select the Power Monitor mode and press
ENTER
.
Zeroing the Power Monitor
Step 4. With no power applied to the DUT, press the
ZERO
soft key from the Power
menu. Wait for a few seconds while the Site Master accumulates data samples of
the quiescent power. When complete,
ZERO ADJ: On
is displayed in the mes-
sage area.
Measuring High Input Power Levels
Step 5. Insert an attenuator between the DUT and the RF detector to protect the Site
Master so that the input power level is less than or equal to 20 dBm.
Step 6. Press the
OFFSET
soft key.
Step 7. Enter the attenuation in dB using the keypad. Press the
ENTER
key to complete
the entry. The message area will show
REL: ON
along with the entered value in
dB.
Displaying Power in dBm and Watts
Step 8. Press the
UNITS
soft key to display power in Watts.
6-1
6