Owner’s Record
The model and serial numbers are located at the rear. 
Record these numbers in the spaces provided below. 
Refer to these numbers whenever you call upon your 
Sony dealer regarding this product.
Model No. ____________________
Serial No. ____________________
WARNING
To reduce the risk of fire or electric shock, do 
not expose this apparatus to rain or moisture.
To avoid electrical shock, do not open the 
cabinet.  Refer servicing to qualified personnel 
only.
WARNING
THIS APPARATUS MUST BE EARTHED.
To disconnect the main power, unplug the AC plug.
For the customers in Canada
This unit has been certified according to Standard CSA 
C22.2 No.601.1
For the customers in the U.S.A.
This equipment has been tested and found to comply 
with the limits for a Class A digital device, pursuant to 
Part 15 of the FCC Rules.  These limits are designed to 
provide reasonable protection against harmful 
interference when the equipment is operated in a 
commercial environment.  This equipment generates, 
uses, and can radiate radio frequency energy and, if not 
installed and used in accordance with the instruction 
manual, may cause harmful interference to radio 
communications.  Operation of this equipment in a 
residential area is likely to cause harmful interference in 
which case the user will be required to correct the 
interference at his own expense.
You are cautioned that any changes or modifications not 
expressly approved in this manual could void your 
authority to operate this equipment.
All interface cables used to connect peripherals must be 
shielded in order to comply with the limits for a digital 
device pursuant to Subpart B of Part 15 of FCC Rules.
WARNING:
Using this unit at a voltage other than 120 V may require 
the use of a different line cord or attachment plug, or 
both.  To reduce the risk of fire or electric shock, refer 
servicing to qualified service personnel.
For the customers in the U.S.A and Canada
When you use this product connected to 240 V single 
phase, be sure to connect this product to a center tapped 
circuit.
Important safeguards/notices for use in the 
medical environments
1. All the equipments connected to this unit shall be 
certified according to Standard IEC60601-1, 
IEC60950-1, IEC60065 or other IEC/ISO Standards 
applicable to the equipments.
2. Furthermore all configurations shall comply with the 
system standard IEC60601-1-1.
Everybody who connects additional equipment to 
the signal input part or signal output part configures 
a medical system, and is therefore, responsible that 
the system complies with the requirements of the 
system standard IEC60601-1-1.
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected 
to other equipment.
4. For this particular equipment, all accessory 
equipment connected as noted above, must be 
connected to mains via an additional isolation 
transformer conforming with the construction 
requirements of IEC60601-1 and providing at least 
Basic Insulation.
5. This equipment generates, uses, and can radiate radio 
frequency energy.  If it is not installed and used in 
accordance with the instruction manual, it may cause 
interference to other equipment. If this unit causes 
interference (which can be determined by 
unplugging the power cord from the unit), try these 
measures: Relocate the unit with respect to the 
susceptible equipment. Plug this unit and the 
susceptible equipment into different branch circuit.
Consult your dealer. (According to standard 
EN60601-1-2 and CISPR11, Class B, Group 1)
6. Model LMD-3250MD is a monitor intended for use 
in a medical environment to display pictures from 
cameras or other systems, other than diagnostic X-
ray equipment.
WARNING
When installing the unit, incorporate a readily 
accessible disconnect device in the fixed wiring, or 
connect the power plug to an easily accessible socket-
outlet near the unit.  If a fault should occur during 
operation of the unit, operate the disconnect device to 
switch the power supply off, or disconnect the power 
plug.