Important safeguards/notices for use in the 
medical environments
1. All the equipment connected to this unit shall be 
certified according to Standard IEC60601-1, 
IEC60950-1, IEC60065 or other IEC/ISO Standards 
applicable to the equipment.
2. When this unit is used together with other equipment 
in the patient area*, the equipment shall be either 
powered by an isolation transformer or connected via 
an additional protective earth terminal to system 
ground unless it is certified according to Standard 
IEC60601-1 and IEC60601-1-1.
3. The leakage current could increase when connected 
to other equipment.
4. The operator should take precautions to avoid 
touching the rear panel input and output circuitry and 
the patient at the same time.
5. Model LMD-1420MD is a monitor intended for use 
in a medical environment to display pictures from 
cameras or other systems. These models are 
nonpatient care equipment with respect to the 
requirement of leakage current.
Caution
When you dispose of the unit or accessories, you must 
obey the law in the relative area or country and the 
regulation in the relative hospital.
WARNING on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) / 
Appliance Connector / Plug with earthing-contacts 
that conforms to the safety regulations of each 
country if applicable.
2. Use the Power Cord (3-core main lead) / Appliance 
Connector / Plug conforming to the proper ratings 
(Voltage, Ampere).
If you have questions on the use of the above Power 
Cord / Appliance Connector / Plug, please consult a 
qualified service personnel.
WARNING on power connection for 
medical use
Please use the following power supply cord.
With connectors (plug or female) and cord types other 
than those indicated in this table, use the power supply 
cord that is approved for use in your area.
*Note: Grounding reliability can only be achieved when the equip-
ment is connected to an equivalent receptacle marked ‘Hospital Only’ 
or ‘Hospital Grade’.
Symbols on the unit
R1.5m
* Patient Area
United States
Canada
Plug Type
HOSPITAL GRADE*
HOSPITAL GRADE*
Female end
E62405, E35708
LR53182, LL022442, 
LL088408
Cord type
E159216, E35496
Min.Type SJT
Min.18 AWG
LL112007-1, LL20262, 
LL32121, LL84494
Min.Type SJT
Min.18AWG
Minimum cord set 
rating
10A/125V
10A/125V
Safety approval
UL Listed
CSA
Symbol
Location
This symbol indicates
Front
Main power switch.
Rear
The equipotential 
terminal which brings the 
various parts of a system 
to the same potential.
Rear
Functional earth terminal
-
Front
Key inhibit
The setting are locked so 
that they cannot be 
changed.