Fluke PM6685R User Manual

Page of 276
used if the command is only four charac-
ters.
If the last character in the command is a
digit, then this digit is appended to the
shortform.
Examples:
Longform
Shortform
:MEASURE
:MEAS
:NEGATIVE
:NEG
:DREGISTER0
:DREG0
:EXTERNAL4
:EXT4
The shortform is always printed in CAPI-
TALS in this manual: :MEASure, :NEG-
ative, :DREGister0, :EXTernal4 etc.
n
Example Language
Small examples are given at various
places in the text. These examples are not
in BASIC or C, nor are they written for
any specific controller. They only contain
the characters you should send to the
counter and the responses that you should
read with the controller.
Example:
SEND→ MEAS:FREQ?
This means that you should program the
controller so that it addresses the counter
and outputs this string on the GPIB.
READ← 1.234567890E6
This means that you should program the
controller so that it can receive this data
from the GPIB, then address the counter
and read the data.
Setting Up the
Instrument
Setting the GPIB Address
The address switches on the rear panel of
the counter are set to 10 when it is deliv-
ered. The address used is displayed when
the instrument is turned on.
If you want to use another bus address,
you can set these switches to any address
between 0 and 30 as shown in the follow-
ing table.
Address
Switch
Settings Address
Switch
Settings
0
00000
16
10000
1
00001
17
10001
2
00010
18
10010
3
00011
19
10011
4
00100
20
10100
5
00101
21
10101
6
00110
22
10110
7
00111
23
10111
8
01000
24
11000
9
01001
25
11001
10
01010
26
11010
11
01011
27
11011
12
01100
28
11100
13
01101
29
11101
14
01110
30
11110
15
01111
1-4 Setting Up the Instrument
Getting Started