Fluke PM6685R User Manual

Page of 276
How to Measure Fast
Speed Summary 7-13
Block Synchronized Measurements
(:ARM:STARt:LAYer:COUNt 1 and TRIG:COUNt <N>)
Speed Improvement Actions
Dead Time Between Measurements
CAL:INT
:AUT
O
OFF
Display
OFF
Realtime
calculations
OFF
([SENS]:INT:FORM
PACK)
Low
resolution
([SENS]:ACQ:RES
LOW)
PM6680B
PM6681
PM6685
9 ms
4.5 ms
9 ms
U
2 ms
1.3 ms
2.5 ms
U
U
0.5 ms
0.12 ms
0.6 ms
U
U
U
0.5 ms
0.025 ms*
0.6 ms
+
1: The GPIB format command will not affect the dead time for the block synchro-
nized mode because the counter captures all data before transferring it to the con-
troller.
+
2: Switching the real time calculations on/off In the block synchronized mode will
significantly decrease the dead time; however, the time for calculations ( 2 ms for
PM6680B/85 and 1 ms for PM6681) is added to the transfer time.
*
In PM6681, low resolution is used for Back-to-Back period measurements.
The measuring time has no effect in this mode. Only the Timestamps are used.