Command Reference 9-17
:CALCulate :LIMit :FAIL?
PM6680B/81/85
Limit Fail
Returns a 1 if the limit testing has failed (the measurement result has passed the
limit), and a 0 if the limit testing has passed.
The following events reset the fail flag:
Power-on
*RST
A :CALC:LIM:STAT OFF ® :CALC:LIM:STAT 8 ON transition
Reading a 1 with this command.
Returned format:
1| 0¿
Example:
SEND® SENS:FUNC 8 FREQ;:CALC:LIM:STAT 8 ON;:CALC:LIM 8 :HIGH8
1E3;READ?;*WAI;:CALC:LIM:FAIL?
READ¬ 1
if frequency ia above 1kHz, otherwize 0
Complies to standards:
SCPI 1991.0, confirmed.
:CALCulate :LIMit :LOWer
PM6680B/81/85
8 «<Decimal data>|MAX|MIN»
Set Low Limit
Sets the value of the ‘Lower Limit’ , i.e., the lowest measurement result allowed be-
fore the counter generates a 1 that can be read with :CALCulate:LIMit:FAIL?,
or by reading the corresponding status byte.
Parameters
Parameter range: –9.9*10
+37
to +9.9*10
+37
.
Returned format:
< Decimal data>¿
*RST condition:
0
Complies to standards:
SCPI 1991.0, confirmed.