Fluke PM6685R User Manual

Page of 276
9-26 Command Reference
:CONFigure :<Measuring Function>
PM6680B/81/85
[8 <parameters>[,(<channels>)]]
Configure the counter for a single measurement
Use the configure command instead of the measure query when you want to
change other settings, for instance, the input settings before making the measure-
ment and fetching the result.
The :CONFigure command controls the settings of the Input, Sense and Trigger sub-
systems in the counter in order to make the best possible measurement. It also
switches off any calculations with :CALC:STATE 8 OFF.
:READ? or :INITiate;:FETCh? will make the measurement and read the resulting
measured value.
Since you may not know exactly what settings the counter has chosen to configure
itself for the measurement, send an *RST before doing other manual set up mea-
surements.
Parameters
<Measuring Function>, <Parameters> and <Channels> are defined on page 9-54.
The optional parameter :SCALar means that one measurement is to be done.
Returned format:
<String>¿
<String> contains the current measuring function and channel. The response is a
<String data element> containing the same answer as for [:SENSe]:FUNC-
tion?.
Example:
SEND® :CONF:FREQ:RAT8(@3),(@1)
Configures the counter for freq. ratio C/A.
See also:
‘Explanations of the Measuring Functions’ starting on page 9-59.
Complies to standards:
SCPI 1991.0, confirmed.