Fluke PM6685R User Manual
9-32 Command Reference
:DISPlay :ENABle
PM6680B/81/85
8 < Boolean >
Display State
Turns On/Off the updating of the entire display section. This can be used for secu-
rity reasons or to improve the GPIB speed, since the display does not need to be
updated. Turning off the display reduces the dead time between measurements
by about 7 ms.
When the display is turned off, the information about the measurement resolution
When the display is turned off, the information about the measurement resolution
is lost. That is, the counter will always send a full 12 digit mantissa independent of
the measurement resolution.
Parameters:
Where <Boolean> = (1 / ON | 0 / OFF)
Returned format:
1|0
¿
*RST condition:
ON
See also:
Chapter 6, ‘How to Measure Fast’.
Complies to standards:
SCPI 1991.0, confirmed.