9-32 Command Reference
:DISPlay :ENABle
PM6680B/81/85
8 < Boolean >
Display State
Turns On/Off the updating of the entire display section. This can be used for secu-
rity reasons or to improve the GPIB speed, since the display does not need to be
updated. Turning off the display reduces the dead time between measurements
by about 7 ms.
When the display is turned off, the information about the measurement resolution
is lost. That is, the counter will always send a full 12 digit mantissa independent of
the measurement resolution.
Parameters:
Where <Boolean> = (1 / ON | 0 / OFF)
Returned format:
1|0
¿
*RST condition:
ON
See also:
Chapter 6, ‘How to Measure Fast’.
Complies to standards:
SCPI 1991.0, confirmed.