9-116 Command Reference
:TRIGger:COUNt
PM6680B/81/85
8 «<Numeric value> | MIN | MAX»
No. of Triggerings on each Ext Arm start
Sets how many measurements the instrument should make for each ARM:STARt
condition, (block arming).
These measurements are done without any additional arming conditions before
These measurements are done without any additional arming conditions before
the measurement. This also means that stop arming is disabled for the measure-
ments inside a block.
+
The actual number of measurements made on each INIT equals to:
(:ARM:START:COUN)*(:TRIG:START:COUNT)
Parameters:
<Numeric value> is a number between 1 and 65535.
MAX gives 65535
MIN gives 1
Example:
SEND® :TRIG:COUN 8 50
Returned format:
<Numeric value>¿
*RST condition:
1
Complies to standards:
SCPI 1991.0, confirmed.