Fluke PM6685R User Manual

Page of 276
Output Subsystem
The analog output is turned off as a de-
fault. You turn it on/off and set the scal-
ing factor under ANALOG OUT in the
aux menu.
Scaling Factor
The scaling factor has two functions:
Its exponent selects which digits to output
on the analog output.
Its value sets what reading should represent
full scale.
As default, the scaling factor is 1 (1E0).
This means that the full scale value is
0.999 and the analog output converts the
fraction (digits to the right of the decimal
point) to a voltage.
The scaling factor should be:
Scaling factor
full scale value
=
1
where full scale value is the value for
which you want the analog output to out-
put its maximum voltage (5 V).
Example:
Take a measurement result, for instance:
12.34567890 E+6 Hz
Represent this result without exponent:
12345678.90 Hz
Multiply this value with the scaling factor,
for instance 0.001.
12345.67890
Take the fractional part of the result:
.67890
Using the Subsystems
Output Subsystem 6-11
5 0 Ω/ 1 M Ω
5 0 Ω/ 1 M Ω
1 2 V r m s - 5 0 Ω
3 5 0 V p - 1 M Ω
5 0 Ω

# 8
5 V   r a n g e
. 6 7 8 9 0   o n   t h e   d i s p l a y  
g i v e s   3 . 3 8 V   w i t h  
s c a l i n g   f a c t o r   1
A N A L O G   O U T  
c o n n e c t o r
A n a l o g   O u t   =   O N ,   s e t   s c a l i n g   f a c t o r   a n d   e x p o n e n t  
s c a l i n g   f a c t o r   1
s c a l i n g   f a c t o r   4
S c a l i n g   f a c t o r   s e l e c t s  
f u l l   s c a l e   v a l u e
S c a l i n g   e x p o n e n t   m o v e s  
i n s e r t i o n   p o i n t
Figure 6-3
The analog output func-
tion.