Fluke PM6685R User Manual

Page of 276
n
Operation Status Group
This group reports the status of the
CNT-8X measurement cycle.
Bit 8 (weight 256) — Measurement
Stopped (MSP)
This bit shows that the counter is not mea-
suring. It is set when the measurement, or
sequence of measurements, stops.
Bit 6 (weight 64) — Wait for Bus Arm-
ing (WFA)
This bit shows that the counter is waiting
for bus arming in the arm state of the trig-
ger model.
Bit 5 (weight 32) — Waiting for Trigger
and/or External Arming (WFT)
This bit shows when the counter is ready
to start a new measurement via the trigger
control option (488.2), for the shortest
possible trigger delay. The counter is now
in the wait for the trigger state of the trig-
ger model.
Bit 4 (weight 14) — Measurement
Started (MST)
This bit shows that the counter is measur-
ing. It is set when the measurement or se-
quence of measurements start.
n
Summary, Operation Status
Reporting
:STAT:OPER:ENAB
Enable reporting of Operation Status in
the status byte.
*SRE 128
Enable SRQ when operation status has
someting to report.
:STAT:OPER?
Reading and clearing the event register of
the Operation Status Register structure
:STAT:OPER:COND?
Reading the condition register of the Op-
eration Status Register structure.
Using the Subsystems
Status Subsystem 6-23
6
8
5 4
0
O p e r a t i o n   S t a t u s   G r o u p
S T A T u s : O P E R a t i o n : C O N D i t i o n ?
S T A T u s : O P E R a t i o n : E V E N t ?
M S P
W F T M S T
W F A
1 5
1 6
3 2
2 5 6
6 4
M e a s u r e m e n t   s t o p p e d
W a i t i n g   f o r   a r m i n g
W a i t i n g   f o r   t r i g g e r i n g
M e a s u r e m e n t   s t a r t e d
Figure 6-11 Bits in the Opeation Sta-
tus Register.