Fluke PM6685 User Manual

Page of 276
Questionable Data/Signal
Status Group
This group reports when the output data
from the CNT-8X may not be trusted.
Bit 14 (weight 16384) — Unexpected
Parameter (UEP)
This bit shows that CNT-8X has received
a parameter that it cannot execute, al-
though the parameter is valid according
to SCPI. This means that when this bit is
true, the instrument has not performed a
measurement exactly as requested.
Bit 10 (weight 1024) — Timeout for
Measurement (TIO)
The counter sets this bit true when it
abandons the measurement because the
internal timeout has elapsed, or no signal
has been detected.
See
also
:SYST:TOUT
and
:SYST:SDET.
Bit 8 (weight 256) Overflow (OFL)
The counter sets this bit true when it can-
not complete the measurement due to
overflow.
n
Summary, Questionable
Data/Signal Status Reporting
:STAT:QUES:ENAB <bit mask>
Enable
reporting
of
Questionable
data/signal status in the status byte.
*SRE 8
Enable SRQ when data/signal is ques-
tionable.
:STAT:QUES?
Reading and clearing the event register of
the Questionable data/signal Register
structure.
:STAT:QUES:COND?
Reading the condition register of Ques-
tionable data/signal Register structure.
6-24 Status Subsystem
Using the Subsystems
O F L
U E P
Q u e s t i o n a b l e   D a t a / S i g n a l   S t a t u s  
G r o u p
T I O
1 5
8
0
1 4
1 0
S T A T : Q U E S t : C O N D ?  
S T A T : Q U E S ?
2 5 6
U n e x p e c t e d  
p a r a m e t e r
T i m e   o u t   f o r   m e a s u r e m e n t
O v e r f l o w
1 6 3 8 4
1 0 2 4
Figure 6-12 Bits in Questionable data
register.