Fluke PM6681 User Manual

Page of 276
9-72 Command Reference
:MEASure :TOTalize :GATed?
PM6680B PM6681
[8 (@«1|2|4|5|6») [,(@«1|2|4|5|6»)]]
Totalize X gated by Y
The counter totalizes the pulses on the primary channel. The totalizing starts when
the gate signal on the secondary channel goes on and stops when the gate signal
goes to off. The polarity of on/off is controlled via the :INPut:SLOPe command of
the gate signal.
Select if the counter should count positive or negative transitions with the
:INPut:SLOPe command of the measuring channel.
Parameters
The first <(@«1|2|4|5|6»)> is the channel to measure on, the second one is the gate channel:
(@1) means input A
(@2) means input B
(@4) means input E (rear panel arming input)
(@5) means input A prescaled by 2
(@6) means the internal reference
If you omit the channels, the instrument measures on input A with input B as the
gate channel.
:MEASure :TOTalize :SSTop?
PM6680B PM6681
[8 ,(@«1|2|4|5|6») [,(@«1|2|4|5|6»)]]
Totalize X start/stop by Y
The counter totalizes the pulses on the primary channel. The totalizing starts when
the gate signal on the secondary channel goes on and stops the next time the gate
signal goes on. The polarity of ON is controlled via the :INPut:SLOPe command
of the Start /stop channel.
Select if the counter should count positive or negative transitions with the :IN-
Put:SLOPe command of the measuring channel.
Parameters:
The first <(@«1|2|4|5|6»)> is the channel to measure on, and the second one is the start/stop
channel:
(@1) means input A
(@2) means input B
(@4) means input E (rear panel arming input)
(@5) means input A prescaled by 2
(@6) means the internal reference
If you omit the channels, the instrument measures on input A with input B is the
start/stop channel.