Anritsu 681XXC User Manual

Page of 241
3-16
REFERENCE
OSCILLATOR
CALIBRATION
The reference oscillator calibration function lets you calibrate the
internal 100 MHz crystal reference oscillator of the 681XXC using an
external 10 MHz, 0 to +10 dBm reference signal.
NOTE
Before beginning calibration, always let
the 681XXC warm up for a minimum of
120 hours.
To perform calibration of the internal reference os-
cillator, first connect the external 10 MHz reference
signal to the 681XXC rear panel
10 MHz REF IN
con-
nector.
Next, press the
SYSTEM
main menu key. At the
System Menu display, press
Cal Menu
to go to the
Calibration Menu (below).
Press
Refrnce Cal
to begin calibration.
Press
Previous Menu
to return to the System Menu
display.
When
Refrnce Cal
is pressed, the Calibration
Status Menu (below) is displayed.
Press
Proceed
to start the calibration.
Press
Abort
to cancel the calibration and return to
the Calibration Menu display.
3-92
681XXC OM
LOCAL (FRONT
REFERENCE
PANEL) OPERATION
OSCILLATOR CALIBRATION
NOTE
This calibration is not applicable to
units having Option16, High Sta-
bility Time Base, installed.