Anritsu 681XXC User Manual
1-9
RECOMMENDED TEST
EQUIPMENT
EQUIPMENT
Table 1-2 lists the recommended test equipment for performing the Se-
ries 681XXC Synthesized Signal Generator operation verification tests
in Chapter 5.
ries 681XXC Synthesized Signal Generator operation verification tests
in Chapter 5.
1-8
681XXC OM
Instrument
Critical Specification
Recommended Manufacturer/Model
Frequency
Counter,
with
Cable Kit
and
External Mixer
Counter,
with
Cable Kit
and
External Mixer
Range: 0.01 to 65 GHz
Input Z: 50
Input Z: 50
W
Resolution: 1 Hz
Other: External Time Base
Input
Other: External Time Base
Input
EIP Microwave, Inc. Models 538B,
548B, or 578B,
with
Cable Kit: Option 590
and
External Mixer:
548B, or 578B,
with
Cable Kit: Option 590
and
External Mixer:
Option 91 (26.5 to 40 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)
Power Meter,
with
Power
Sensors
with
Power
Sensors
Range: –30 to +20 dBm
(1
(1
m
W to 100 mW)
Anritsu Model ML2437A or ML2438A,
with
Power Sensors:
with
Power Sensors:
MA2474A (0.01 to 40 GHz)
MA2475A (0.01 to 50 GHz)
MA2475A (0.01 to 50 GHz)
Oscilloscope
Bandwidth: DC to 150 MHz
Vertical Sensitivity: 2 mV/
division
Horiz Sensitivity: 50 ns/
division
Vertical Sensitivity: 2 mV/
division
Horiz Sensitivity: 50 ns/
division
Tektronix, Inc. Model TAS485
Table 1-2.
Recommended Test Equipment
GENERAL
RECOMMENDED
INFORMATION
TEST EQUIPMENT