Anritsu 683XXB User Manual
1-12
RECOMMENDED TEST
EQUIPMENT
EQUIPMENT
Table 1-2 provides a list of recommended test equipment needed for
the performance verification, calibration, and troubleshooting proce-
dures presented in this manual.
the performance verification, calibration, and troubleshooting proce-
dures presented in this manual.
1-12
682XXB/683XXB MM
INSTRUMENT
CRITICAL
SPECIFICATION
RECOMMENDED
MANUFACTURER/MODEL
USAGE
(1)
Spectrum Analyzer,
with
External Mixers
and
Diplexer Assy
with
External Mixers
and
Diplexer Assy
Frequency Range: 0.01 to 65 GHz
Resolution Bandwidth: 10 Hz
Resolution Bandwidth: 10 Hz
Tektronix, Model 2794, with
External Mixers:
External Mixers:
WM780K (18 to 26.5 GHz)
WM780A (26.5 to 40 GHz)
WM780U (40 to 60 GHz)
WM780E (60 to 90 GHz)
WM780A (26.5 to 40 GHz)
WM780U (40 to 60 GHz)
WM780E (60 to 90 GHz)
Diplexer Assy: 015-0385-00
C, P, T
Spectrum Analyzer
Frequency Range: 20 Hz to 40 MHz
Resolution Bandwidth: 3 Hz
Resolution Bandwidth: 3 Hz
Hewlett-Packard, Model 3585B
P
Frequency Counter
with
Cable Kit
and
External Mixer
with
Cable Kit
and
External Mixer
Frequency Range: 0.01 to 65 GHz
Input Impedance: 50
Input Impedance: 50
W
Resolution: 1 Hz
Other: External Time Base Input
Other: External Time Base Input
EIP Microwave, Inc. Models 538B, 548B,
or 578B, with
Cable Kit: Option 590
and
External Mixer:
or 578B, with
Cable Kit: Option 590
and
External Mixer:
Option 91 (26.5 to 40 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)
C, P
Power Meter,
with Power Sensor
with Power Sensor
Power Range: –30 to +20 dBm
(1
m
W to 100mW)
Hewlett-Packard Model 437B, with
Power Sensor:
Power Sensor:
HP 8487A (0.01 to 50 GHz)
C, P
Power Meter,
with Power Sensor
with Power Sensor
Power Range: –30 to +20 dBm
(1
m
W to 100mW)
ANRITSU ML4803A, with
Power Sensor:
Power Sensor:
MP716A4 (50 to 75 GHz)
C, P
Digital Multimeter
Resolution: 4-1/2 digits (to 20V)
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M
W
AC Accuracy: 0.07% +100 counts
(to 20 kHz)
AC Input Impedance: 1 M
W
John Fluke, Inc., Model 8840A, with
Option 8840A-09K (True RMS AC)
Option 8840A-09K (True RMS AC)
C, T
Frequency Reference
Frequency: 10 MHz
Accuracy: 5 x 10
Accuracy: 5 x 10
-12
parts/day
Absolute Time Corp., Model 300
P
Frequency Standard
Frequency: 10 MHz
Accuracy: 1
Accuracy: 1
´
10
–10
parts/day
Spectracom Corp., Model 8161
C
Function Generator
Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Hewlett-Packard, Model 8116A
C
Function Generator
Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Functions: 0.1 Hz to 100 kHz sine and
square waveforms
Hewlett-Packard, Model 33120A
C
Table 1-2.
Recommended Test Equipment (1 of 3)
GENERAL
RECOMMENDED
INFORMATION
TEST EQUIPMENT