Tektronix AFG3000 User Manual

Page of 230
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual
3-107
*TST?
This command performs a self-test and returns the results.
Group
Calibration and Diagnostic
Related Commands
DIAGnostic[:ALL]?
Syntax
*TST?
Arguments
None
Returns
<NR1>
where
<NR1>=0 indicates that the self-test completed without errors.
<NR1>
≠0 indicates that the arbitrary/function generator detected an error.
Examples
*TST?
performs a self-test and returns 0 if no error is reported.
*WAI (No Query Form)
This command prevents the instrument from executing further commands or 
queries until all pending commands that generate an OPC message are complete.
Group
Synchronization
Related Commands
*OPC
Syntax
*WAI
Arguments
None
Examples
*WAI
prevents the instrument from executing any further commands or queries until all 
pending commands that generate an OPC message are complete.
NOTE. 
The self-test can take several minutes to complete. During this time, the 
arbitrary/function generator does not execute any commands. Do not power off the 
instrument during the self-test.